How to calibrate the CS200

The following details the calibration procedure for the Intercomp CS200 crane scale.
There are four parameters that can be set before entering the actual calibration. When
changing a value, advance at least to “save

save

save

save” before turning off so that your changes are

saved.

Step

Description

Note

Recommended

rd rt

rd rt

rd rt

rd rt Read rate

1 = 1 reads/sec
2 = 2 reads/sec
3 = 4 reads/sec
4 = 7 reads/sec
5 = 10 reads/sec
6 = 18 reads/sec
7 = 25 reads/sec

4

AZT

AZT

AZT

AZT

Auto Zero Tracking

0 = disabled, 1,2,3

3

dely

dely

dely

dely calibration delay

0 = no delay
1 = 7 second delay for cal
points

1

grad

grad

grad

grad graduation

4 = 5 lb
5 = 2 lb
6 = 1 lb
7 = 0.5 lb
8 = 0.2 lb
9 = 0.1 lb
10 = 0.05 lb
11 = 0.02 lb
12 = 0.01 lb

cap

cap

cap

cap

capacity

save

cal

cal

cal

cal

calibration type

1 =1 point cal linear
0101 = 3 point cal

0101

LL00

LL00

LL00

LL00 Zero read

LL01

LL01

LL01

LL01 Cal load 1

Enter load 1

LL02

LL02

LL02

LL02 Cal load 2

Enter load 2

LL03

LL03

LL03

LL03 Cal load 3

Enter load 3

Start up.

1. Turn the power OFF.
2. Remove the calibration blocking shunt to allow entry into calibration mode.
3. Turn the power ON.
4. Wait for scale to warm up (3 minutes from power on).
5. Press ZERO and BACKLIGHT together and release to enter the calibration

mode.
The scale shows “rd rt

rd rt

rd rt

rd rt”нннн. The read rate is the number of internal reads per

second the scale performs. A higher setting will result in a faster responding