Weight calibration – Rice Lake Intercomp CS 200 - Users Manual User Manual

Page 13

Advertising
background image

CS200, Users

August, 2009

13

Weight Calibration

10. The scale shows “CAL

CAL

CAL

CAL”. Press ZERO and backlight keys together and release.

The recommended calibration type is 3 point linear. Enter “0101”. Otherwise
enter “1” for a 1 point calibration.

11. The scale shows “LL00

LL00

LL00

LL00”. With no weight on the pad, press ZERO and

BACKLIGHT together and release. This reads the pad zero.

12. The scale shows “LL01

LL01

LL01

LL01”. Apply the first weight/force. Press the ZERO and

BACKLIGHT keys together and release. Then enter the value of the weight you
just applied. Press ZERO and BACKLIGHT together and release to enter this
number.

13. The scale shows “LL02

LL02

LL02

LL02”. Apply the second weight. Press the ZERO and

BACKLIGHT keys together and release. Then enter the value of the weight you
just applied. Press ZERO and BACKLIGHT together and release to enter this
number.

14. The scale shows “LL03

LL03

LL03

LL03”. Apply the third weight. Press the ZERO and

BACKLIGHT keys together and release. Then enter the value of the weight you
just applied. Press ZERO and BACKLIGHT together and release to enter this
number.

15. Return the calibration blocking shunt to it's original location to prevent accidental

entry into the calibration mode.

16. Verify the calibration.
17. Calibration complete.

Advertising