Weight calibration – Rice Lake Intercomp CS 200 - Users Manual User Manual
Page 13
CS200, Users
August, 2009
13
Weight Calibration
10. The scale shows “CAL
CAL
CAL
CAL”. Press ZERO and backlight keys together and release.
The recommended calibration type is 3 point linear. Enter “0101”. Otherwise
enter “1” for a 1 point calibration.
11. The scale shows “LL00
LL00
LL00
LL00”. With no weight on the pad, press ZERO and
BACKLIGHT together and release. This reads the pad zero.
12. The scale shows “LL01
LL01
LL01
LL01”. Apply the first weight/force. Press the ZERO and
BACKLIGHT keys together and release. Then enter the value of the weight you
just applied. Press ZERO and BACKLIGHT together and release to enter this
number.
13. The scale shows “LL02
LL02
LL02
LL02”. Apply the second weight. Press the ZERO and
BACKLIGHT keys together and release. Then enter the value of the weight you
just applied. Press ZERO and BACKLIGHT together and release to enter this
number.
14. The scale shows “LL03
LL03
LL03
LL03”. Apply the third weight. Press the ZERO and
BACKLIGHT keys together and release. Then enter the value of the weight you
just applied. Press ZERO and BACKLIGHT together and release to enter this
number.
15. Return the calibration blocking shunt to it's original location to prevent accidental
entry into the calibration mode.
16. Verify the calibration.
17. Calibration complete.