How to calibrate the pw800 – Rice Lake Lift Truck/Pallet Jack Scales - Intercomp PW800 User Manual

Page 12

Advertising
background image

PW800, Users

Rev O, January, 2008

12

How to Calibrate the PW800

The following details the calibration procedure for the PW800 scale. There are eight
parameters that can be set without moving the calibration blocking switch, followed by five more
parameters and calibration that require the calibration blocking switch be in the 'CAL' position.

Step

Parameter

Note

Default

ee-ee

ee-ee

ee-ee

ee-ee Skip

0,1=EE-07

EE-07

EE-07

EE-07, 2=LL-00

LL-00

LL-00

LL-00

0

00

0

EE-00

EE-00

EE-00

EE-00 Sample Rate

1 to 127

8

88

8

EE-01

EE-01

EE-01

EE-01 Sleep Display Time

0 to 120; 0=off

20

20

20

20

EE-02

EE-02

EE-02

EE-02 Auto-off Time

0 to 255; 0=off

120

120

120

120

EE-03

EE-03

EE-03

EE-03 Baud Rate

0 to 7

0

00

0

EE-04

EE-04

EE-04

EE-04 Power up in KG

0=lb, 1=kg

0

00

0

EE-05

EE-05

EE-05

EE-05 Decimal point position for

APW

0=none, 1=tenths,
2=hundreths

0

00

0

EE-06

EE-06

EE-06

EE-06 Serial Output

0=continuous, 1=on demand

1

11

1

Save information

Check for calibration blocking switch

EE-07

EE-07

EE-07

EE-07 AZT

0=off,1=0.6,2=1,3=3

2

22

2

ee-08

ee-08

ee-08

ee-08 Zero Range

0=off,1=on

0

00

0

EE-09

EE-09

EE-09

EE-09 Canadian Specifications

0=off,1=on

0

00

0

EE-10

EE-10

EE-10

EE-10 Initial Zero Range

0=off,1=on

0

00

0

EE-11

EE-11

EE-11

EE-11 Graduation

0 to 11

6

66

6

EE-12

EE-12

EE-12

EE-12 Reserved

EE-13

EE-13

EE-13

EE-13 Circuit Board ID #

1 to 159

1

11

1

Save

LL-00

LL-00

LL-00

LL-00 Zero read

Enter capacity

LL-01

LL-01

LL-01

LL-01 First weight

Enter first weight

LL-02

LL-02

LL-02

LL-02 Second weight

Enter second weight

LL-03

LL-03

LL-03

LL-03 Third weight

Enter third weight

Save

Start up
1. Remove the screws around the front panel.
2. Move the calibration blocking switch (located on back of the circuit board) to the

'CAL' position if you intend to calibrate.

3. Turn scale power ON.
4. Wait for scale to warm up (3 minutes from power on).
5. Using the UNITS button, set the scale to the appropriate units that you intend to

calibrate with.

6. Press the NET/GROSS and MODE switches simultaneously. This enters you into

the calibration routine.


First seven parameters
7. The scale shows “EE-EE

EE-EE

EE-EE

EE-EE”. Press the NET/GROSS and MODE keys. To skip to

“EE-07

EE-07

EE-07

EE-07” enter a ‘1’. To skip to “LL-00

LL-00

LL-00

LL-00” enter ‘2’. No skips will occur with an entry of

‘0’.