PW800, Users

Rev O, January, 2008

15

2: The IZSM (initial zero setting mechanism on power up) must be within +/- 10%

of the zero obtained at calibration.

3: The push-button zero and AZSM can only operate within +/- 2% of the IZSM.

4: The over-capacity point is 103% of capacity above the IZSM.

18. The scale shows “EE-10

EE-10

EE-10

EE-10”. Press the NET/GROSS and MODE keys. The scale

shows the current 'Initial zero range' setting. Use the following table to select the
'Initial zero range' setting:

Setting

Initial zero range

0

Off

1

On

A '0' setting implies: The initial zero setting mechanism (IZSM) will work over the
entire range of the scale capacity and the over-capacity point is 103% above the
zero obtained at calibration.
A '1' setting implies: The IZSM must be within +/- 10% of the zero obtained at
calibration and the over-capacity point is 103% above the IZSM.


19. The scale shows “EE-11

EE-11

EE-11

EE-11”. Press the NET/GROSS and MODE keys. The scale

shows the current graduation selection. Use the following table to select a
graduation value.

Settings

Count by in lb

Count by in kg

0

100

50

1

50

20

2

20

10

3

10

5

4

5

2

5

2

1

6

1

0.5

7

0.5

0.2

8

0.2

0.1

9

0.1

0.05

10

0.05

0.02

11

0.02

0.01


20. The display shows “EE-12

EE-12

EE-12

EE-12”. This parameter is reserved for future use. Press

NET/GROSS and MODE keys together twice to continue.


21. The scale shows “EE-13

EE-13

EE-13

EE-13”. Press the NET/GROSS and MODE keys. The display

shows the current circuit board ID#. This is used in multi scale applications. The
circuit board ID is used for identification.