Maintenance

9

If hooks or chains were used during calibration, remove these and the test weights from the scale. With

all weight removed, press

to rezero the scale. This function adjusts the zero and span calibration

values. The indicator displays

*CAL*

while the zero and span calibrations are adjusted. When

complete, the adjusted A/D count for the zero calibration is displayed. Press

to enter the value, then

press to return to the CALIBR menu.

7. Press until the display reads

CONFIG

, then press to exit setup mode.

Note

To perform the calibration using EDP commands or Revolution

®

, consult the indicator operator's manual.