To check a load bar separately, it must be completely disconnected from the scale's electrical circuit. This is

accomplished by disconnecting the load bar at the scale's junction box, Figure 6-1. However if the junction box

contains a Weights & Measures seal, consult your scale dealer before proceeding. To disconnect a load bar, loosen

the screws on the circuit board for the load bar in question, and remove the wires. Resistance measurements can

now be taken as shown in Table 6-1, and should correspond with the values in the "One Load Bar" column in the

same way as described above. If not, consult your scale dealer. The dealer may ask you to check the individual load

cells.

Wire from
Load Bar 1

Black (-EX)

Green (+SIG)

Red (+EX)

White (-SIG)

Shield

Black (-EX)

Green (+SIG)

Red (+EX)

White (-SIG)

Shield

Wire from
Load Bar 2

Black (-EX)

Gr

een (+SIG)

Red (+EX)

White (-SIG)

Shield

Wire to Instrument

Figure 6-1. Junction Box Wire

When checking an individual load cell, it must be completely disconnected from the scale's electrical circuit. To

accomplish this, the wires will need to be taken apart at the load bar cable splice. See Figure 6-2 for the wiring of a

load bar system. The circuit diagram for the Load Cells of a single load bar is shown in Figure 6-4. Remember that

each load cell contains five wires, but each load bar has only four wires running to the junction box. All five wires

must be disconnected to separate that load cell from the circuit.