3.2

VIRTUI Configuration

VIRTUi

can be used to set

iQube

configuration parameters. When configuration is complete, data is downloaded

to the

iQube

.

To configure the

iQube

using

VIRTUi

:

1. With the PC and

iQube

connected and the

VIRTUi

program running, select

iQube

Configuration

from the

Settings

menu.

Figure 3-10.

VIRTUi Tracer AVi Display

2. Select the

Cells

display. Use this display to select the load cell used in the

Tracer AVi

system by checking

the box to the left of the cell. Enter load cell data such as factory sensitivity, serial number, and capacity
on the General Information sheet for the cell.

3. Select the

Platforms

display. The cell that was checked on the Cells screen is now listed as Available

Load Cell for the

iQube

.

4. Click on

Platform 1

.

5. Double-click on the load cell to move the cell into the Assigned Load Cell column for Platform 1.

As the cell is added to the Assigned Cell column, a Section Format Diagram is displayed at the bottom of
the screen.

6. Select the

Systems

display.

7. Click on

System 1

.

8. Platform 1 is now shown in the Assigned Platforms field. Click the

Platform 1

box to assign Platform 1 to

System 1. Use the General display to enter scale system parameters.

3.2.1

Downloading to the Tracer AVi

Once configuration is complete, you must download the configuration data from the PC to the

iQube

.

1. Select

Send Configuration

to write the values to the

iQube

.

2. Click

Begin

to initiate the download. Downloading may take up to 30 seconds.