0 calibration, Calibration – Rice Lake Tracer AVi - v2.0 User Manual

Page 17

Advertising
background image

Calibration

13

4.0

Calibration

VIRTUi Calibration

With the

iQube

connected to a PC running the

VIRTUi

program, do the following:

1. From the

Settings

menu, select

Calibration

then

Multi-Point.

Figure 4-1. Platform Selection

2. Remove all weight from the scale platform. Click on

Calibrate

. The word

Transmitting

is displayed while

the zero value is calculated. This process can last up to 45 seconds.

Figure 4-2. Calibration

3. Enter the test weight value in the

Calibration Weight

field. You are now ready to measure the load cell

output using the Multi-Point procedure.