2 patient + chair – Rice Lake Wireless Wheelchair Scale D300 Series User Manual

Page 11

Advertising
background image

Note

D300 Series Wireless Wheelchair Scale Installation & Operation Manual — Using the Scale

7

Once the handheld controller is turned off, the chair weight is gone.

5. Attach mini ramps back onto the scale pads.
6. Remove wheelchair off of scale pads.

Enter Chair Weight

To manually enter a chair weight, press the down arrow to highlight

Enter Chair Weight

.

1. Prior to manually entering the wheelchair weight, place the included wheel stops and mini ramps onto the

scale pad as shown in Figure 4-1.

2. Roll the wheelchair backwards onto the center of each pad.
3. Set the brakes on the wheelchair and remove mini ramps prior to weighing.
4. Press

OK

.

5. Press the up arrow or the down arrow to begin the numbers scrolling in 5 lb increments (press and hold for

faster scrolling) until you get to the chairs’ correct weight and press

OK

. The controller should read

Is This

Correct?

Yes

should be highlighted and if the weight is correct, press

OK

. If the weight is not correct, press

the down arrow to highlight

NO

and press

OK

. Repeat until correct.

6. Attach mini ramps back onto the scale pads.
7. Remove wheelchair off of scale pads.

4.2.2

Patient + Chair

Use the down arrow to the highlighted

Patient + Chair

. Use the following steps to find the weight of the patient and

the chair.

1. Place each scale pad on a solid, flat surface (tile, cement, hardwood).
2. Place the wheel stops on the scale pads.
3. Press the down arrow key and scroll down until

Zero Out Pads

is highlighted and press

OK

.

4.

Check and remove anything from the scale pads

is displayed. Press

continue

to initiate the zero out function.

(when the controller is in process the screen will go dim for approximately five seconds and while the

controller is zeroing out each pad, the scale pads will illuminate green.

Important

If any of the four pad indicator lights are red or yellow in color, or if the controller indicates

COMM ERROR, TRY AGAIN

, press the up arrow to highlight the

Yes

prompt. Press

OK

and repeat the

zeroing out process.

5. After the controller completes the process, the display shows

Zeroing Complete

and will verify that all four

pad numbers on the controller display are indicated with green circles.

6. The controller will ask

Repeat Zero Out.

If all numbers are circled in green in the controller display, then

highlight

NO

and then press

OK

and return back to the main menu screen.

7. Place mini ramps in front of each scale pad making sure the scale pads are placed behind the wheelchair.
8. Roll the wheelchair onto the scale pads, as close to the center of the pad as possible and ensure brakes are

on.

9. Assist patient into the chair.
10. Remove mini ramps prior to weighing the patient.
11. Press

Patient + Chair

and the display will show

Weighing

. During this time the display screen will dim for

approximately five seconds and the scale pads will be lit green. Patient weight will then be displayed. The

display

Save?

is shown. Press

Yes

to save that weight and will be saved until the next use. If the user does

not wish to save the weight, press the down arrow to highlight

NO

and then press

OK

to return back to the

main menu.

12. Attach mini ramps back onto scale pads.
13. Remove wheelchair and patient off of scale pads.

Advertising