Taring the beam – Seca 745 band_1 User Manual

Page 36

Advertising
background image

36 •

Balancing the beam

1. Ensure that both sliding weights are in the “0”

position.

2. Undo the knurled screw on the tare weight.

3. Move the tare weight to balance the beam.

4. Wait until the beam has stopped swinging.

NOTICE!

Incorrect measurement due to force shunt

If you touch the scale, weight will not be
measured correctly.
– Ensure that you do not touch the scale while

the beam is swinging.

5. Check whether the beam is balanced.

Pointer and counterpointer must be at exactly or
approximately the same level.

6. Repeat Step 3 if required.

7. Fix the tare weight in position using the knurled

screw.

8. Turn the fine-tare screw until the pointer and

counterpointer are exactly opposite one another.
The scale is operational.