2 weighing, Performing a weighing operation – Seca 745 band_1 User Manual

Page 37

Advertising
background image

Operation • 37

English

6.2

Weighing

CAUTION!

Risk of falling, faulty operation

To prevent falls from the scale and faulty opera-
tion, you must ensure before each use that the
scale is operational.
– Ensure that the tray is correctly fixed in

position (see “Put on tray” on page 34).

– Perform a function check (see “Function

check” on page 40).

– Check whether the beam is balanced (see

“Checking the position of the beam” on
page 35).

Performing a weighing

operation

1. Ensure that the locking lever of the beam is re-

leased (see “Releasing/locking the beam” on
page 35).

2. Place the baby on the tray.

WARNING!

Risk of falling

Baby scales generally stand on raised work sur-
faces. If the baby falls from this surface, this
may result in serious irreversible or fatal injuries.
– Ensure that the device is steady and level.

– Route connecting cables (if present) in such a

way that people cannot trip over them.

– Never leave a baby unsupervised.

3. Move the large sliding weight until the beam is

balanced approximately.
Pointer and counterpointer must be at
approximately the same level.

4. Wait until the beam has stopped swinging.

NOTICE!

Incorrect measurement due to force shunt

If you touch the scale, weight will not be
measured correctly.
– Ensure that you do not touch the scale while

the beam is swinging.

3.

Advertising
This manual is related to the following products: