Welch Allyn MicroTymp 3 portable tympanometric instrument - User Manual User Manual

Page 46

Advertising
background image

46

Printing 226-Hz results

Welch Allyn MicroTymp3 portable tympanometric instrument

Print interpretive messages (Switch #2)

Print tympanogram only or tympanogram and data (Switch #3)

Manufacturing switch (Switch #4)

No messages

Depress the ON portion of the switch to select this option.
This causes messages which interpret the tympanogram to
not be included on the printout.

Refer to the Printout formats section on

page 42

for more

information on these messages.

Messages

Depress the OFF portion of the switch to select this option.
This causes the messages which interpret the
tympanograms to be included on the printout.

OFF

1

2

3

4

1

2

3

4

OFF

Tympanogram only

Depress the ON portion of the switch to select this option.
Only the tympanogram and the GR (Width) numeric value will
print.

Tympanogram and data Depress the OFF portion of the switch to select this option.

Both the tympanogram and its corresponding numeric data
will print.

1

2

3

4

OFF

1

2

3

4

OFF

This switch is used during manufacturing only. Leave this
switch in the OFF position. The Printer/Charger will not
operate normally if this switch is on.

1

2

3

4

OFF

Advertising