7.

Tighten screw in clockwise direction. To avoid stripping the screw threads, do not
tighten excessively.

8.

Place the Handle into the charging station for a minimum of 10 seconds to enable the
battery protection circuit.

Recharge the battery

To recharge the battery, place the Handle in the Printer/Charger well with the LCD and
buttons facing you.

The 3.7V Lithium-ion battery used in the Handle, when fully charged, provides a full day of
operation without the need for recharging and yields a minimum of 300 double-ear tests.
This makes the instrument optimal for mass screening or off-site situations where there
may not be a need to print, but there is a need for continuous operation.

A fully drained battery should be recharged overnight (16 hours).

Note

The unit will not power up if you do not enable the battery protection circuit.

Micr

oT

ymp

®

3

P

o

rt

a

ble

T

ym

pan

o

me

tr

ic

In

st

ru

m

e

nt

0297

IEC TYPE 3
Skaneateles Falls, NY USA

US Pat #5,383,097

ANSI TYPE 4

Patent Pending

C

US

74227

23650

REF

Caution To prevent equipment damage, charge the Handle only with the
Printer/Charger

(#7117x).

Note

The Handle may be charged indefinitely without damage to the battery.

Slight heating of the Handle during charging is normal.

The battery will self-discharge gradually over a period of approximately 60
days when stored at room temperature (70°F/21°C); storage at higher
temperatures accelerates the discharge rate.