Documenting patient turn and resetting counter – Welch Allyn EarlySense System 1.0 - Equivalent to model EverOn 1.0 - User Manual User Manual
Page 52
46
Figure 29: Patient Turn Box – Disabled Counter
Figure 30: Patient Turn Box – Enabled Counter
Documenting Patient Turn and Resetting Counter
After turning the patient, press Patient Turned within 1 minute from the time that the
patient turn was performed. The system logs the event (Figure 31). In addition, the
system detects the corresponding change in body movement and verifies the event. The
verification will appear in the printed report of the same day (See Reports, page 59)
Figure 31: Recording Patient Turned Event
If the patient turn interval passes without the Patient Turned button being pressed by the
user, a Patient Turn alert will occur (Figure 32) indicating that the counter has exceeded
its time threshold. The alert message will disappear by pressing Ok or by documenting a
patient turn by pressing Patient Turned. If Ok was pressed, but the Patient Turned
button was not pressed within 15 minutes, the alert will go off again.