Welch Allyn Connex ProBP 3400 digital blood pressure device - User Manual User Manual
Page 20
WARNING NIBP measurements may be inaccurate in the presence of excessive
motion artifact. Minimize extremity and cuff motion during blood pressure readings.
WARNING The position and physiologic condition of the subject can affect a
blood pressure reading.
WARNING If the blood pressure cuff is not at heart level, note the difference in
reading due to the hydrostatic effect. Add the value of .2 kPa (1.80 mmHg) to the
displayed reading for every 2.5 cm (inch) above heart level. Subtract the value of .2
kPa (1.80 mmHg) from the displayed reading for every 2.5 cm (inch) below heart
level.
WARNING Proper blood pressure cuff size and placement is essential to the
accuracy of the blood pressure determination. See Blood pressure cuff selection
for sizing information.
14
NIBP warnings and cautions
Welch Allyn Connex® ProBP™ 3400 digital blood pressure device