Electromagnetic immunity

The device is intended for use in the electromagnetic environment specified below. The customer or the user

of the device should assure that it is used in such an environment.

Immunity test

IEC 60601 test

level

Compliance level Electromagnetic environment -

guidance

Electrostatic

discharge (ESD)
IEC 61000-4-2

±6 kV contact
± 8 kV air

± 6 kV contact
± 8 kV air

Floors should be wood, concrete or

ceramic tile. If floors are covered with

synthetic material, the relative humidity

should be at least 30%.

Electrical fast

transient/burst
IEC 61000-4-4

±2 kV for power

supply lines
±1 kV for input/

output lines

±2 kV for power

supply lines
±1 kV for input/

output lines

Mains power quality should be that of a

typical commercial or hospital

environment.

Surge
IEC 61000-4-5

± 1 kV differential

mode
± 2 kV common

mode

± 1 kV differential

mode
± 2 kV common

mode

Mains power quality should be that of a

typical commercial or hospital

environment.

Voltage dips, short

interruptions and

voltage variations on

power supply input

lines
IEC 61000-4-11

>95% dip in 0.5 cycle
60% dip in 5 cycles
30% dip for 25 cycles
>95% dip in 5

seconds

>95% dip in 0.5 cycle
60% dip in 5 cycles
30% dip for 25 cycles
>95% dip in 5

seconds

Mains power quality should be that of a

typical commercial or hospital

environment. If the user of the device

requires continued operation during power

mains interruptions, it is recommended

that the device be powered from an

uninterruptible power supply or a battery.

Electromagnetic immunity

The device is intended for use in the electromagnetic environment specified below. The customer or the user

of the device should assure that it is used in such an environment.

Immunity test

IEC 60601 test

level

Compliancelevel Electromagnetic environment -

guidance

Portable and mobile RF communications

equipment should be used no closer to any

part of the device, including cables, than

the recommended separation distance

calculated from the equation applicable to

the frequency of the transmitter.

Recommended separation distance

Conducted RF
IEC 61000-4-6

3 Vrms
150 kHz to 80 MHz

3 Vrms

d= (1.17)

Radiated RF
IEC 61000-4-3

3 V/m 80 MHz to 1

GHz

3 V/m

d= (1.17)

80 to 800 MHz

d= (2.33)

800 MHz to 2,5 GHz

58 Guidance and manufacturer's declaration