Inspecting the equipment, Testing the electrocardiograph, Replacing the battery – Welch Allyn CP 50 Plus 12-lead resting electrocardiograph - User Manual User Manual

Page 36

Advertising
background image

3.

Dampen a cloth with any of the acceptable cleaning solutions, and wipe the exterior

of the patient cable, reusable electrodes, and electrocardiograph. Dry all

components with a clean, soft cloth, or paper towel.

WARNING The patient cable and reusable elctrodes should be
cleaned between each patient.

4.

Before you turn on the electrocardiograph again, wait at least 10 minutes for all

traces of liquid to evaporate.

Inspecting the equipment

Perform the following inspections daily.

Check for cracked or broken patient cable, patient electrodes, power cord,

communications cables, display, and enclosure.

Check for bent or missing pins on all cables.

Check all cable and cord connections; reseat if any connectors are loose.

Testing the electrocardiograph

Welch Allyn recommends verifying proper operation of the electrocardiograph once a

year to ensure reliability. See Verifying proper operation.

Whenever the electrocardiograph is serviced or whenever problems are suspected,

verify continued electrical safety of the device using IEC 60601-1 or ANSI/AAMI ES1

methods and limits.

WARNING Only qualified service personnel should perform leakage
current tests.

Test for the following:

Patient leakage current

Chassis leakage current

Earth leakage current

Dielectric strength (AC and patient circuits)

Replacing the battery

Replace the battery under these circumstances:

It loses its charge quickly.

You have charged it, and the electrocardiograph still does not turn on when

unplugged.

The first time that you press the power button after installing a new battery, the

electrocardiograph goes through some diagnostic tests that cause it to take longer than

usual to power up.

Discard the old battery appropriately. Contact your local authorities concerning recycling.

32 Maintenance

CP 50™ and CP 50 Plus™ 12-lead resting electrocardiograph