Hydrolysis step-by-step procedure using the

ANKOM

HCl

Hydrolysis System

To perform hydrolysis on prepared samples, follow the procedure detailed below.

1. Turn the instrument power switch ON.

2. Turn the cold water supply on.

3. Pull the Lid Latch up to open the Vessel Lid.

4. Place the Multi-bag Holder into the Hydrolysis

Vessel.

5. Pour 500ml of 3N HCl into the Hydrolysis

Vessel to cover the ports in the back.

6. Roll up a piece of paper towel and place it on

the Vessel Lid just above the Hinge as shown
to the right. This will help prevent drips of
liquid from getting on the Hinge after a run.

7. Push the Lid Latch down to close the Vessel Lid.

8. Press ENTER on the Keypad.

9. Use the UP/DOWN arrows on the Keypad to enter the hydrolysis time (suggested time is 60 minutes).

10. Press ENTER on the Keypad.

11. Use the UP/DOWN arrows on the Keypad to enter the hydrolysis temperature (suggested temperature is

90°C).

12. Press ENTER on the Keypad.

13. Use the UP/DOWN arrows on the Keypad to enter the rinse time (suggested time is 20 minutes).

14. Press ENTER on the Keypad.

15. Make sure the Vessel Lid is closed (a message on the Display will remind you to do this).

Display

Keypad

Lid Latch

Vessel Lid

Hinge on
Vessel Lid

Hydrolysis
Vessel

Rolled-up
Paper Towel

Ports in back
of Vessel