I. introduction, I.1 components – Brookfield LFRA Texture Analyzer User Manual

Page 3

Advertising
background image

Brookfield Engineering Labs., Inc.

Page 3

Manual No. M/05-370-B0507

I. INTRODUCTION

The Brookfield LFRA Texture Analyzer measures the resistance of a material to an applied force.

The force is applied through a vertical compression of the sample and the resistance is measured

through a load cell. This resistance can be interpreted to provide a measure of the texture or other

physical characteristics of the material.

The LFRA Texture Analyzer is the most recent development of the original Boucher Jelly Tester.

This original product was created as the Stevens Texturemeter evolving into the CNS Farnell LFRA

TA (Leatherhead Food Research Association Texture Analyzer). Brookfield’s enhanced design

maintains the characteristics of the original equipment while expanding the measurement capabili-

ties and ease of use by incorporating modern digital technology.

The principle of operation of the LFRA Texture Analyzer is to drive a probe into a material. The

resistance of the material to this compression is measured by a calibrated load cell and reported

in units of grams on the instrument display. The measured load will be a function of the probe

geometry, speed of compression, test method, and sample preparation.

There are four load cell ranges available for the LFRA Texture Analyzer offered by Brookfield:

Model

Load Cell Range (grams)

LFRA-100

100g

LFRA-1000

1000g

LFRA-1500

1500g

LFRA-4500

4500g

Higher load cell ranges are used for testing materials with a stronger structure.

I. Components

Please check to be sure that you have received all components, and that there is no damage. If

you are missing any parts, please notify Brookfield Engineering or your local Brookfield agent

immediately. Any shipping damage must be reported to the carrier.

Part.Number.

Quantity

Component

Part Number

Quantity

LFRA Texture Analyzer

varies

1

Probe

varies

per order

Operating Manual

M05-370

1

Power Cord

1

115 Volt or

DVP-65

1

230 Volt

DVP-66

Optional

Bloom Strip

TA-CS-100 or

1

TA-CS-1000

Calibration Weight Set (1)

TA-CW-100C, -1000C, -1500C, -4500C

1

Standard Probe kit

TA-P-KIT2

1

TexturePro Lite Software

TA-T-PRO

1

Note: 1. Optional components may include probes or fixtures. See Appendix A for details.

2. See III.9 for Calibration Checking (pg. 14).

Advertising