I.5 installation – Brookfield LFRA Texture Analyzer User Manual

Page 5

Advertising
background image

Brookfield Engineering Labs., Inc.

Page 5

Manual No. M/05-370-B0507

Position:

Range: 0-75 mm

Resolution: 0.1mm

Accuracy: 0.1mm

Output:

Computer: RS232 Compatible Serial Port

Environmental 0°C to 40°C temperature range (32°F to 104°F)

Conditions:

20% - 80% relative humidity, non-condensing atmosphere

I.5 Installation

1) Prepare a clean, level surface.

NOTE: This instrument is a sensitive load measuring device. It should be installed on a

clean, solid and level bench surface which is free from external vibrations.

2) Unpack and remove the LFRA Texture Analyzer from the shipping container.

The LFRA Texture Analyzer weighs 15 kg (33 pounds). Use caution when lifting the
unit out of the packaging.

3) Place the LFRA Texture Analyzer on a sturdy, level surface. Adjust the four feet to ensure

that the instrument is stable.

4) Remove any additional components from the shipping package. Save the shipping container

and packaging for future use.

5) Make sure that the AC power switch at the rear of the LFRA Texture Analyzer is in the OFF

position. Connect the power cord to the socket on the back panel of the instrument and plug

it into the appropriate AC line. Position instrument so that the power cord can be removed

easily.

The AC input voltage and frequency must be within the appropriate range as shown
on the model and serial tag of the instrument (located on the back of the LFRA).

The LFRA Texture Analyzer must be earth grounded to ensure against electronic fail-
ure!

6) If appropriate, connect interconnecting cable which is supplied with TexturePro Lite software

to the serial port for connection to a computer.

7) Turn the power switch to the ON position. The startup screen will indicate the firmware

version and load range of the LFRA (Figure II.1).

LFRA VERSION X.X
TEXTURE ANALYZER

1000 GRAM UNIT

INITIALIZING

Figure I.1