Appendix c - warranty and repair service 30, Appendix b - troubleshooting – Brookfield LFRA Texture Analyzer User Manual

Page 29

Advertising
background image

Brookfield Engineering Labs., Inc.

Page 29

Manual No. M/05-370-A1205

Appendix B - Troubleshooting

If no power to the instrument:
Check:

• User replaceable fuses: two (2) fuses, 4 amp, 5 x 20mm, time-lag

• Fuses located within power entry module on the rear of the instrument.

• If fuses blow continually, serious damage to the instrument could result. Contact repair

services immediately.

Probe won’t attach:

• Check thread on probe and probe shaft - remove any dirt

Base table won’t move:

• Ensure locking knob is loose

• Ensure fine adjust nut is loose

Base table will not travel to the lowest position:

• Loosen fine adjust nut and rotate base table

Instrument is unsteady:

• Ensure that bench top is stable

• Adjust the LFRA feet

Probe doesn’t move when start is pressed:

• Ensure emergency stop button is not pressed in (rotate clockwise)

• Ensure test mode is not Static Load

• Ensure probe is not in contact with sample (or base table)