I.2 Utilities

Input Voltage:

115 VAC or 230 VAC

Input Frequency:

50/60 Hz

Power Consumption:

150 VA

Fuse:

Two 4 amp, 5 x 20mm, Time-lag

Power Cord Color Code:

United States

Outside United States

Hot (live)

Black

Brown

Neutral

White

Blue

Ground (earth)

Green

Green/Yellow

Main supply voltage fluctuations are not to exceed

±

10% of the nominal supply voltage.

I.3 Units

The LFRA Texture Analyzer uses the SI system of units for all parameters.

Parameter

Unit

Abbreviation

Load

Grams

g

Distance

Millimeter

mm

Time

Seconds

s

Speed

Millimeter per second

mm/s

I.4 Specifications

Load:

Range

Resolution

Model

Grams

Grams

Accuracy*

LFRA-100

0-100

0.01

±

0.5%FSR

LFRA-1000

0-1000

0.10

±

0.5%FSR

LFRA-1500

0-1500

0.20

±

0.5%FSR

LFRA-4500

0-4500

0.50

±

0.5%FSR

FSR = Full Scale Range
*This value applies to the complete range of operating temperatures as defined
in Section I.4. This value can be improved if the LFRA is controlled to a single
temperature.

Trigger Point:

Range

Resolution

Model

Grams

Grams

LFRA-100

0.1-10

0.01

LFRA-1000

0.2-100

0.10

LFRA-1500

0.2-150

0.10

LFRA-4500

0.6-500

0.10

Speed:

0.1 to 1.0 mm/s in increments of 0.1 mm/s
1 to 10 mm/s in increments of 1 mm/s
Accuracy:

±

0.1% of setpoint