Brookfield UL Adapter, Enhanced User Manual

Page 3

Advertising
background image

Brookfield Engineering Laboratories Inc.

-  -

Operating Instructions M06-084-A0507

4. Level the viscometer. General operating procedures for making viscosity measurements are described in the

viscometer operating instruction manual.

5. UL Adapter spindle factors are shown on page 4. The factor is used to calculate viscosity when using Dial

Reading Viscometers. All Brookfield digital model Viscometers and Rheometers calculate the viscosity value

automatically. The spindle entry code for the UL Adapter is 00.

Notes

• Maximum recommended viscosity for measurement with the UL Adapter is 2000 cP (mPa•s). If viscosity

exceeds 2000 cP, the material being measured may be too viscous to immerse the spindle into the sample

chamber; damage to internal parts of the viscometer may result.

• For tubing and fluid recommendations when using a Constant Temperature Bath, refer to the table below:

Fluid

Recommended

Recommended

Note

Temperature

Fluid

Tubing

-10°C to 15°C TC-FLUID 2

1

ULA-45

R, 2

Do Not Use Gum

Rubber Tubing With

This Fluid

15°C to 65°C

Water

Gum Rubber

or Fluran

R

65°C to 100°C TC-FLUID 3

3

ULA-45

R

Do Not Use Gum

Rubber Tubing With

This Fluid

Table 1

R

Fluran is a Registered Trademark of Norton Co.

1. Use only laboratory grade ethylene glycol. Do not use automobile anti-

freeze which contains materials that can damage the equipment.

2. Fluran tubing (5/16" ID) and clamps are offered in a kit, part # ULA-45A.

3. Do not use high viscosity oil. Recommended is 50 centipoise.

TEMPERATURE

BATH

BATH OUTLET

BATH INLET

VISCOMETER

PIVOT CUP

(on Viscometer)

WATER

JACKET

OUTLET

WATER

JACKET

INLET

TUBE END CAP, ULA-34

SAMPLE CHAMBER

MOUNTING COLLAR

KNOB

Figure 3: ULA-49EAY Water Jacket Connections