Ul adapter factors, Lv models rv-ha-hb models – Brookfield UL Adapter, Enhanced User Manual

Page 4

Advertising
background image

Brookfield Engineering Laboratories Inc.

- 4 -

Operating Instructions M06-084-A0507

The factor is used to calculate viscosity in units of centipoise (cP) or milli Pascal seconds (mPa•s) when using

Dial Reading or Model DV-I Viscometers. The Viscometer Dial Reading/DisplayValue x Factor = cP (mPa•s).

Factors are built-in for Digital Models DV-E, DV-I+, DV-I Prime, DV-II+ (all versions), and DV-III (all versions)

— no calculations required.

UL Adapter Factors

Figure 4: Pivot Cup Compatibility

When an older dial reading visco-
meter is used, the viscometer pivot
cup may need to be replaced with the
current design. If your viscometer
has a Type I or Type II pivot cup, the
cup should be replaced with a Type
III cup. Contact Brookfield or your
Brookfield agent for information.

LV Models

RV-HA-HB Models

Shear

Shear

Factor

Speed

Rate

Speed

Rate

(RPM)

(sec

-1

)

Factor

(RPM)

(sec

-1

)

RV

HA

HB

60.0

73.38

0.1

100.0

122.30

0.64

1.28

5.12

30.0

36.69

0.2

50.0

61.15

1.28

2.56

10.24

12.0

14.68

0.5

20.0

24.46

3.20

6.40

25.60

6.0

7.34

1.0

10.0

12.23

6.40

12.80

3.0

3.67

2.0

5.0

6.12

12.80

25.60

1.5

1.83

4.0

4.0

4.89

16.00

0.6

0.73

10.0

2.5

3.06

25.60

0.3

0.37

20.0

2.0

2.45

1.0

1.22

0.5

0.61

Shear Rate Constant = RPM x .