background image

6 | Operation

BÜCHI Labortechnik AG

36/60

Operation Manual Rotavapor® R­100

u

In applications where no vacuum is applied, lower the temperature of the heating
bath.
That will cause less solvent to evaporate.

6.2.4

Feeding solvent during distillation

Precondition:
R An external vacuum pump is connected and running.
R A feed tubing leading from the glass stopcock to the evaporating flask is fitted.

u

Connect a supply tubing (2) for the solvent to the glass stopcock (1) and hold it in
the solvent.

u

Turn the glass stopcock until the handle is pointing downwards.
Solvent is drawn into the evaporating flask by the vacuum.

u

Make sure that there is not more than 3 kg of solvent in the evaporating flask.

u

Close the glass stopcock.