MODEL CT-

2000 PROBLEM ANALYSIS

A successful CT-2000 self-test is an indicator that there are no detectable
errors and checked components are functional . There are certain condi-
tions, however, that cannot be verified by these automatic tests. These
include:

1. Viscosity of the bath liquid.
2. Temperature probe installation
3. Motor-stirrer functionality
4. Temperature probe functionality
5. Heater operation

Problems with the first three items listed above may cause incorrect
temperature stability. If items 4-5 fail, the bath may be non-operational.

Fault isolation

For assistance in troubleshooting your CT-2000 consult the troubleshoot-
ing guide below. If you are unable to resolve the difficulty, call
CANNON

®

Instrument Company for assistance.

Problem

Possible cause/solution

Bath does not appear to have power.

• Power cable not connected to outlet or rear of

electrical drawer. Attach cables.

• Power cord unplugged. Attach cord to

appropriate outlet (see electrical requirements
on CT-2000 rear panel).

• Power out on mains. Restore power.

Bath illumination not functioning.

• Separate switch for LIGHT must be on.
• Fluorescent lamps may be defective. Replace.
• Lamps may be out of sockets. Tighten lamps.
• Lamp ballast may be defective. Replace.
• Ballast Power Supply may not be working

properly. Replace.

Bath liquid not agitated.

• Stirring motor not functioning. Check connection

of stirring motor on rear panel. Visually check
for working motor by connecting to outlet
with the proper voltage (see motor label for
correct voltage).