Assembly procedure, Glass jar installation – Cannon Instrument CT-2000 (Contant Temperature Bath) User Manual

Page 8

Advertising
background image

6

CANNON

®

CT-2000 Constant Temperature Bath

Revision 1.0g—February, 2012; CANNON

®

Instrument Company

2139 High Tech Road • State College, PA 16803 • USA

Assembly procedure

1. Unpack the bath unit housing

and move it to its permanent
location on a stable labora-
tory bench or table.

2. Remove all eight screws from

the stainless steel top covers
(see Figure 2).

3. Disconnect all external

cabling (probes, heaters, and
float switch) from the upper
rear panel of the CT-2000.

4. Remove the front top cover and rear top cover (see Figures 3 and 4).

Use caution when removing the rear top cover because the tempera-
ture control probes and heating elements are attached to it.

Glass jar installation

1. Remove the glass jar from its box.

WARNING

The glass jar is heavy. Use caution when lifting it.

2. Make sure the rubber

support ring is seated
properly around the
bottom jar opening in the
bath unit (see Figure 5).

3. Place the rubber gasket

around the top rim of the
jar (see Figure 6). The

Figure 5: Seating the support ring

Figure 2: Detaching top cover

Figure 3: Removing
front top cover

Figure 4: Removing
back top cover