Checkline DT-326 User Manual

Page 33

Advertising
background image

30

5

Specifications

Model Number

DT-326

Units of measure

FPM (flashes per minute) or Hz (frequency)

Measuring range

1 - 2,000Hz

60 - 120,000 FPM

Accuracy

+/-0.02% (at 73°F [23°C])

Resolution
of flash rate
setting in
internal
mode

Unit;
Hz

Measuring
range:
1-200 Hz

0.01Hz

Measuring
range:
1-2000 Hz

0.1Hz

Unit;
FPM

Measuring
range:
60 - 12,000 FPM

60.0 - 3,000.0 : 0.1 FPM
3,000.2 - 6,000.0 : 0.2 FPM
6,000.5 - 12,000.0 : 0.5 FPM

Measuring
range:
60 - 120,000
FPM

60 - 30,000 : 1 FPM
30,002 - 60,000 : 2 FPM
60,005 - 100,000 : 5 FPM
100,010 - 120,000 : 10 FPM

Flashing time (Duty)

0.1° - 2.5°/360° setting can be changed by 0.1° inc rements

Mode

Internal

Set and change
flashing rate

Change flash rate by using dial, or “×2” and “1/2” key

Phase Shift

Forward/backward by 3° increments

Unit change

“FPM” or ”Hz”

External
trigger

Flash rate
(Input frequency)

1 – 166 Hz
60 – 10,000 FPM

Input pulse width More than 50 us

Unit change

“FPM” or ”Hz

Parameter
setting

Measuring range 1 - 2000Hz (60 - 120,000 FPM) range or

1 - 200Hz (60 - 12,000 FPM)

Trigger Edge

Positive or Negative edge

Delay time/angle Delayed time : 0 - 999ms *1

Delayed angle : 0 - 359° *1

Back light

Selectable- ON or OFF

Power Saving
Function

Flash (Auto off)

Stop flashing automatically after 2 min.

Back light (Auto off)

Turn off after 2 minutes of inactivity.

Power (Auto off)

Power off unit after approx. 5 minutes in stand-by mode

Input/Output (I/F)

External Trigger Input

Photo Coupler Input, Current 12mA (approx.)

External Trigger output

Photo Coupler (Open collector) Output
Voltage: DC30V or less, Current: 30mA or less,
Pulse width: approx.200us

Sensor Power Supply

DC12V/max.40mA

*1

Built-in delay time = 60us

Advertising