11.0 CALIBRATION


By a zero point adjustment you can correct a too high deviation at the calibration
check.

For that please proceed as follows:

Press the left key twice to reach the menu level. Navigate
to the menu item „Options“ using the arrow keys. Confirm
by pressing the button.

Select the menu item „Calibrate“ using the arrow keys
and confirm by pressing .




This menu item has to be unlocked. For that you have to enter the 4-digit serial
number of your humimeter PM4. You can find the serial number on the top right
edge of the display a short time after switching on the instrument, or on the sticker in
the battery slot.

Press the third button as long as the black bar stands on
the first digit of the serial number. After a waiting time of a
few seconds the device accepts the entered digit. Now
the other 3 digits have to be entered. When the 4-digit
number has been entered correctly (see picture), press
the button to confirm.

Hold the humimeter PM4 in one hand up in the air and
make sure that nothing stands behind the steel clamps
(distance min. 200 mm). Now press the second button
( ) with one finger.

The calibration procedure will take a few seconds. During
that time, the instrument has to be held up in the air and
the measuring field (steel clamps) must NOT be touched.

The device is ready when the shown picture on the right
disappears automatically.


Electromatic Equipment Co., Inc

-9-

12.0 CHARCHING THE BATTERIES

Connect the provided USB cable to the device and the
other end of the cable to a PC or another USB charging
adapter. It takes about 6 hours to charge the completely
discharged batteries. Please make sure that the
temperature during the charging process is between
0°C and 45°C
, as otherwise the batteries may be
destroyed.

Hardware Reset

In case that your humimeter device does not respond to any
key press or cannot be switched on, there can be carried
out a hardware reset directly by the customer. Please note:
the accumulator must be charged before your start the
following procedure:
Slide a pair of tweezers or a small screw driver below the
cover cap. If your tool stick, try again at another position
(the arrows on the image show the positions). Do not use
any force. Open the cover cap and push the push button
using your tool. Now the humimeter will restart. Fix the
cover cap again.
If it will not restart, please contact your technical support.







13.0 EXEMPTION FROM LIABILITY


For miss-readings and wrong measurements and of this resulting damage we refuse
any liability. This is a device for quick determination of moisture. The moisture
depends on multiple conditions and multiple materials. Therefore we recommend a
plausibility check of the measuring results. Each device includes a serial number
and the guarantee stamp. If those are broken, no claims for guarantee can be
made. In case of a faulty device, please contact Checkline Europe.