MOST COMMON REASONS FOR MEASURING ERRORS



Product temperature beyond ambient conditions: The temperature of the

measuring device and of the product should be nearly equal.

Wrong switch position

Uneven pressing

Wrong application on rolls

Press your measuring device lengthwise to the longitudinal axis of the
product roll, otherwise the bearing area of the sensor is too small.

Product pile not thick enough

Electrically conductive material

All metallic objects as well as electrically conductive packaging film or paper
coloured with carbon black influence the measuring result in a negative way
and must no be positioned within a range of 500 mm below the sensor area.

Exceeding of measuring range:

This can occur depending on the type of product and the position of the
selector switch. It is indicated on the display by blinking and means that the
measuring accuracy is less precise.

ELECTROMATIC Equipment Co., Inc.

-11-

P2 Examination table

Copy this table and fill in the displayed values – without temperature correction –
and the ambient temperature. Enclose a copy of the data if the instrument should
be adjusted again by the factory.

serial number measuring instrument:_____________

serial number proof plate:______________

temperature :______°C

date :_____________

switch-

position

set

measuring

value

displayed

value

P2

temperature-

compensation

value

drift

in

%

readjust

tolerance

S1 6.4

0.6

S2 5.7

0.6

S3 5.1

0.5

S4 4.5

0.5

S5 3.8

0.4

S6 3.1

0.4



P4 Examination table

Copy this table and fill in the displayed values – without temperature correction –
and the ambient temperature. Enclose a copy of the data if the instrument should
be adjusted again by the factory.

serial number measuring instrument:____________

serial number proof plate:_____________

temperature :______°C

date :_____________

switch-

position

set

measuring

value

displayed

value

P4

temperature-

compensation

value

drift

in

%

readjust

tolerance

S1 4,3

0,6

S2 4

0,6

S3 3,7

0,5

S4 3,3

0,5

S5 3

0,4

S6 2,6

0,4

LM5 Examination table