9 Technical specifications

37

9 Technical specifications

9.1

The NucleoCounter

Specificity

The NucleoCounter counts mammalian cell nuclei stained
with the DNA specific fluorescent dye, propidium iodide.

Sample consumption

100 µl is recommended for NucleoCassette loading.
Approximately 60 µl is loaded into the cassette.

Analysis volume

Approximately 2 µl of the sample loaded into the
NucleoCassette is transported to the chamber and is
analyzed.

Measurement range

The measurement range is 5x10

3

– 2x10

6

cells/ml.

The optimal measurement range is 1x10

5

– 2x10

6

cells/ml.

Analysis time

When pressing “Run” on the NucleoCounter the result will
be displayed within 30 seconds.

Operation

Menu-controlled by means of keyboard and LCD display.

Physical data

Weight
Height
Width
Depth

3 kg
26 cm
38 cm
22 cm

Power

See Section 10.2 Power Supply

Input
Fuse

12VDC (11-13VDC)
UL Listed fuse:
600mA/250Vac (UL 248)

Power consumption
NucleoCounter

Peak
Ready mode
Standby

25 W
2.5 W
2 mW

Operation conditions

Maximum relative humidity 80 percent for temperatures up
to 31°C decreasing linearly to 65 percent relative humidity
at maximum 35°C; minimum temperature 15°C.

USB

USB, version 1.1.
Note: Does not support USB Hubs.

Printer

Thermal or Impact printer with RS-232 interface and cable
supplied by ChemoMetec A/S.

Other environmental operating conditions for the NucleoCounter

Indoor use only.

Altitude

Up to 2000 m.

Mains supply

Refer to the specifications for the Power supplies in section
10.2 Power supply.