12 Appendix 2: Extended NucleoCounter

®

NC-100

TM

Settings

47

12 Appendix 2: Extended NucleoCounter

®

NC-100

T M

Settings

Enabling of extended instrument sensitivity

Enabling of extended instrument sensitivity

Enabling of extended instrument sensitivity

Enabling of extended instrument sensitivity (F566

(F566

(F566

(F5661)

1)

1)

1)

8

88

8

The extended instrument sensitivity is activated by the user function F5661
(“F5661 + ↵”) on the keypad. Once you have activated it, it cannot

cannot

cannot

cannot be deactivated!

Instrument Sensitivity (F11)

Instrument Sensitivity (F11)

Instrument Sensitivity (F11)

Instrument Sensitivity (F11)

The appropriate setting is selected by the user function F11 (“F11 + ↵”) on the keypad.
Press “1” to toggle up and “2” to toggle down through the menu.

For firm ware 4.10 the options are Type 1 – 7.
For firm ware 4.11 the options are Type 1 – 9.

In case you have changed the setting of the instrument, the display will show “Save
parameters?” when you turn it off. Push the (“•”) key on the keypad so the instrument
stores this setting permanently.

The NucleoCounter NC-100 has 9 different levels of sensitivity, which can be adjusted to
count the different types of cells. Type 1 is the least sensitive (default NC-100) and
should be used for the counting of mammalian cells.
We recommend that NucleoView is used to select the best Type for your cells, by
ensuring that cell signals are separated from debris and other noise from the sample.

8

Available in firm ware 4.10 and forward.