6!NucleoView!SCC0400!!

21!

!

Figure,25,.,Tools,.,Barcode,menu,item,

The!Barcode!function!activates!a!dialogue!as!shown!in!the!following!figure.!By!entering!
sample!identification!in!the!input!field!and!activating!the!Run!button!the!next!sample!will!
get!the!corresponding!identification.!It!is!possible!to!configure!a!barcode!reader!in!such!a!
way!that!it!reads!the!barcode!information!and!upon!completion!enters!Carriage0Return!
(CR)!instruction.!This!allows!the!barcode!reader!to!initiate!the!next!measurement.!

!

Figure,26,.,Barcode,dialogue,

6.2.6 Tools-L-Sample-
The!mixing!of!reagent!and!sample!must!be!done!in!accurate!ratio!in!order!to!assure!the!
analytical!performance!of!the!NucleoCounter!SCC0400.!This!mixing!ratio!can!be!defined!
using!the!Tools%U%Instrument!menu!item.!

!

Figure,27,.,Tools,.,Sample,menu,item,

The!Sample!option!of!the!Tools!menu!can!be!used!to!define!the!exact!dilution!of!the!
sample!mixture.!The!sample!and!reagent!are!intended!to!be!mixed!as!1!volume!milk!and!2!
volumes!reagent!(1+2).!As!it!can!be!convenient!to!determine!the!volume!dispensed!
accurately!rather!than!adjusting!pipettes!or!dispenser!the!Sample!function!allows!the!
operator!to!enter!the!accurate!volumes.!Note%deviation%from%the%intended%mixing%ratio%of%
1+2%will%affect%the%accuracy%of%the%NucleoCounter%SCCU400%system!
!!