O.r – zero tolerance range (set as % of capacity)

o

100

o

2

D – division size

o

50, 20, 10, 5, 2, 1, 0.5, 0.2, 0.1, 0.05, 0.02, 0.01, 0.005, 0.002, 0.001,

0.0005, 0.0002, 0.0001

Filter – (intended to minimize effects of motion, vibration, and wind currents)

Flf – Flush filter. Allows switch to fast filter rate if weight changes quickly

Tare – enable tare function

o

Off – tare inactive

o

On – tare is active

o

On-Clr – automatically clears when gross wt returns to zero.

Hl – overload limit (set as % of capacity)

o

102.5

o

103

o

105

o

110

o

300

Ul – underload limit (set as % of capacity below zero)

o

5.5

o

10

o

25

o

50

o

100

D rate – display update rate (set in tenths of seconds)

o

From 0.1 to 1.0

P1 – serial port 1 output format

o

GTN – if in gross mode will print gross weight. If in net mode, will print

gross/tare/net weights

o

nEt – will print net only

P2 – serial port 2 output format

o

GTN

o

nEt


APP menu

X.xxupd – display of micro volts per division. Should be greater than 1.0. Read only.
If less than 1.0, adjust division size.

Acc – Enable/disable accumulator

A thld – Autoprint and accumulator threshold. Number of divisions to meet or exceed to
trigger auto print or accumulator.