Ionode IJ pH V5.1 User Manual

Page 28

Advertising
background image

Ionode IJ Instruction Manual

28

Symptom Possible

Causes

Remedy

Low Slope
(<94%)

Contaminated
Buffers

Replace buffers

Membrane
not Clean

See 4.6

Troubleshooting

Drift

Membrane
Aged

Allow more time for
stabilisation, or
replace

Electrode
Polarised

Replace electrode

Slow
Response

Membrane
not Clean

See 4.6

Troubleshooting

Drift

Membrane
Aged

Allow more time for
stabilisation, or
replace

Incorrect
Sample
Reading
after
successful
calibration

Ground Loop.
Often Occurs
in Process
Systems

Verify by removing
the sample from its
environment and
measuring in a glass
beaker. May require
special circuitry.

Junction
Blocked or
unclean

See 4.6

Troubleshooting

Drift

Displays
pH7 for all
Buffers

Electrical
Short

Check connector

Internal Short Replace electrode

Displays
pH4-5 for
all Buffers

Membrane or
Stem
Cracked

Replace electrode

Advertising