Metal 6.1. application – ijau, Application – ijag – Ionode IJ pH V5.1 User Manual

Page 41

Advertising
background image

Ionode IJ Instruction Manual

41

6. Metal
6.1. Application – IJAu

The IJAu uses a gold rod as sensor and is
designed for determining ORP measurements
in cyanide solutions, and in other applications
where Pt is easily poisoned.

See 5.3 and 5.3.1 for Cleaning/Maintenance

6.2. Application – IJAg

The IJAg uses a silver rod as sensor and can
be used for determining the activity of silver
ions by direct or indirect potentiometry, and for
determination of halides and cyanide by
argentimetric titration. The most common
application is for determination of chloride or
salt in environmental and food samples by
titration with silver nitrate.

A typical titration curve for potassium chloride
vs 0.1M silver nitrate is shown below.

Maintenance

– clean off excess silver chloride

deposits from the electrode after use. Replenish
electrolyte (with 1M KNO

3

regularly), and do not

leave the electrode immersed in silver
containing solutions when not in use.

Titration Curve of Potassium Chloride (acidified) vs 0.1M silver nitrate

using IJAg Silver Electrode with 10% KNO3 in sleeve junction

0

50

100

150

200

250

300

350

400

450

500

0.0

5.0

10.0

15.0

20.0

25.0

Volume of 0.1M silver nitrate added

Potential (mV)

Series1