Maintenance, Cleaning – Ionode IJ pH V5.1 User Manual

Page 35

Advertising
background image

Ionode IJ Instruction Manual

35

different potentials and irreversible redox
couples.

4) Use only a high impedance mV meter.

5) Use fresh standards.

6) Use

polyethylene,

Teflon

™ or glass

beakers and stir with a magnetic Teflon

coated stirrer bar.

5.3. Maintenance

1) When not in use, store the electrode in 2-

3M KCl. This assists in maintaining a stable
reference potential and is preferred to
distilled water or an ORP standard.

2) How often does the sleeve electrolyte need

replacing? This depends on the sample
being measured and the desired accuracy.
It is best determined by experience. An
error can be determined by comparing
readings before and after renewal of the
electrolyte on the same sample.

3) For long term storage, remove and clean

the sleeve. Replace the sleeve without
electrolyte and attach the wetting cap filled
with 20% Potassium chloride solution.

5.3.1. Cleaning

Refer to 4.4.1 pH

Cleaning

Metals (e.g. copper and zinc) that have plated
onto the platinum tip can be removed with dilute
(approx 0.1M) nitric acid.

Advertising