Calibration Function

When there is difference between the internal measurement value (R.V.) and the actual
dimension of the object, the value can be corrected. There are two correction methods as
below: One-point calibration and Two-point calibration

One-point calibration
Only tilt (span) is corrected while the zero point (offset) is not corrected.

Two-point calibration
Both the zero point (offset) and the tilt (span) are corrected.

Note

• If the following conditions are not satisfied, the screen below is

displayed when the setting is complete. Make the setting again.

For one-point calibration

For two-point calibration

-99.999

d (offset value A) d99.999

"3-7 Zero Shift Function (Shifting the Internal Measurement Value (R.V.))"
(page 3-16)

• If the measurement mode is set to the "Pin interval judgment mode"

or "Pin diameter judgment mode", the calibration function cannot be
used. Only "Standard" and "User setting" can be switched on the
setting screen.

"1. Measurement mode" (page 4-8)

Correction

R.V. before correction

R.V. [

5'V] you wish to display

Correction

R.V. (2) before correction

R.V. (1) before correction

Offset value A

R.V. (2) [

5'V] after correction

R.V. (1) [

5'V] after correction

'TT

LASER

BANK

0

1

2

3

HI

GO

LO

R.V.

ANALOG

HI

SHIFT

ZERO SHIFT

TIMING

LO

ALIGNMENT

5'6

HOLD

CALC

CHECK

0.5

d (R.V. you wish to display)

(R.V. before correction)

d2

0.5

d (R.V. (2) you wish to display) - (R.V. (1) you wish to display)

(R.V. (2) before correction) - (R.V. (1) before correction)

d2