Area scaling, Hold function, S-h p-h b-h a-h auto.p auto.b auto.a s-h – KEYENCE SK-1000 Series User Manual

Page 12: Sample hold, E sk-1000 im, Zarea scaling correction table

Advertising
background image

12

E SK-1000 IM

5. Area scaling

This function configures the settings related to area scaling. The area scaling
function corrects errors in the measured value due to size differences of
measurement targets by setting a correction value according to the size of the
target.

Refer to the correction table shown below to setting the correction value.

z

Area scaling correction table

(1) High-precision mode

(2) Wide-range mode

[How to use the correction table and setting example]
Take a wide-range mode as an example for setting a correction value where the
distance to the target is 100 mm and the size of target is φ400 mm.
In this case, use the correction table for the wide-range mode.

The 100 mm on the vertical axis and 200 mm and -200 mm on the horizontal axis
make two intersecting points, crossing the lines marked with the S symbol, which
indicate a correction value of 1.1. So, input 1.1 into the amplifier unit.

Item

Setting range

initial value

Area scaling

0.10 to 1.50

1.00

NOTICE

When the measurement target is larger than the outermost lines
of the correction table:

Example: In wide-range mode, with a distance of 100 mm to the
target, and a target size of
φ780 mm, if the target is larger than the

outermost lines indicating the correction value of 1.3, set the
correction value to 1.3.

The same is true when using high-precision mode. If the target is
larger than the outermost lines indicating a correction value of
1.0, set the correction value to 1.0.

5. ScL

HOLD

HI

V

kV

ALARM

BANK

ZERO SHIFT

TIMING

TEMP

%RH

HI

LO

R.V.

0

1

2

3

LO

GO

Press the S/T buttons and specify the correction value.
(Setting range: 0.10 to 1.50)

0

5

10

15

20

25

30

35

40

45

50

-200

-150

-100

-50

0

50

100

150

200

1.0

0.9

0.8

0.7

0.6

1.0

1.0

0.9

0.9

0.8

0.8

0.7

0.7

0.6

0.6

Measurement area [mm]

D

ist

an

ce

to t

he

tar

g

et

[

m

m

]

60

70

80

90

100

110

120

-450

-350

-250

-150

-50

50

150

250

350

450

1.3

1.2

1.1

1

0.9

0.7

1.3

1.3

1.2

1.2

1.1

1.1

0.9

0.9

0.7

0.7

1

1

Measurement area [mm]

Distan

ce

to

the

ta

rge

t [mm]

60

70

80

90

100

110

120

-450

-350

-250

-150

-50

50

150

250

350

450

1.3

1.2

1.1

1

0.9

0.7

1.3

1.3

1.2

1.2

1.1

1.1

0.9

0.9

0.7

0.7

1

1

Measurement area [mm]

Distan

ce

to

the

ta

rge

t [mm]

Correction
value is 1.1.

The measurement area ranging from
-200 mm to +200 mm covers

φ

400 mm.

60

70

80

90

100

110

120

-450

-350

-250

-150

-50

50

150

250

350

450

1.3

1.2

1.1

1

0.9

0.7

1.3

1.2

1.1

1.3

1.2

1.1

0.9

0.7

1

0.9

0.7

1

Measurement area [mm]

D

ista

nc

e t

o th

e

target

[mm]

The measurement area ranging from
-390 mm to +390 mm covers

φ

780 mm.

Correction
value is 1.3.

6. Hold Function

The hold method is set for the judgment value (P.V.).
Functions other than "Auto peak hold", "Auto bottom hold" and "Auto absolute
value hold" require the use of the external input (timing input).

Reference

While the judgment value (P.V.) is held, the hold indicator on the sensor
amplifier lights up.

„

Sample hold

When the timing input is turned ON, the judgment value (P.V.) is held.

z

When "Level" is set for "7. Timing input" (page 14)

The judgment value (P.V.) is held only while the timing input is ON.

z

When "Edge" is set for "7. Timing input" (page 14)

The judgment value (P.V.) is held in between rising edges of the timing input.

Hold function

Description

Default

value

Sample hold

When the timing input is turned ON, the judgment
value (P.V.) is held.
"Sample hold" (page 12)

c

Peak hold

The maximum value of the specified period
(sampling period) is held as the judgment value
(P.V.).
"Peak hold" (page 13)

Bottom hold

The minimum value of the specified period
(sampling period) is held as the judgment value
(P.V.).
"Bottom hold" (page 13)

Absolute value hold

Within the specified period (sampling period), the
absolute value of the maximum measured value is
held as the judgment value (P.V.).
"Absolute value hold" (page 13)

Auto peak hold

Until the hold-reset input is turned ON, the
maximum value is held as the judgment value (P.V.).
External input is not necessary.
"Auto peak hold" (page 13)

Auto bottom hold

Until the hold-reset input is turned ON, the minimum
value is held as the judgment value (P.V.).
External input is not necessary.
"Auto bottom hold" (page 13)

Auto absolute value hold

Until the hold-reset input is turned ON, the absolute
value of the maximum measured value is held as
the judgment value (P.V.).
External input is not necessary.
"Auto absolute value hold" (page 14)

6. HLd

HOLD

HI

V

kV

ALARM

BANK

ZERO SHIFT

TIMING

TEMP

%RH

HI

LO

R.V.

0

1

2

3

LO

GO

[HOLD] ON

S-H

P-H

b-H

A-H

Auto.P

Auto.b

Auto.A

S-H

Display value

Time

Judgment value (P.V.)
Internal measurement value (R.V.)*

1

Timing Input ON

OFF

Display value

Time

Judgment value (P.V.)
Internal measurement value (R.V.)*

2

Timing Input ON

OFF

Power ON or
Reset input *

1

After the power is turned on or reset is input, the judgment value (P.V.) is
[- - - - -] until the timing input is first turned ON.

Advertising