13

E SK-1000 IM

„

Peak hold

The maximum value of the specified period (sampling period) is held as a
judgment value (P.V.).

z

When "Level" is set for "7. Timing input" (page 14)

When the timing input is turned ON, the maximum value of the internal
measurement value (R.V.) during the sampling period is held as the judgment
value (P.V.). When the timing input is turned OFF, the next sampling period begins.

z

When "Edge" is set for "7. Timing input" (page 14)

When the timing input is turned ON, the maximum value of the internal
measurement value (R.V.) during the sampling period is held as the judgment
value (P.V.). Also, when the timing input is turned ON, the next sampling period
begins.

„

Bottom hold

The minimum value of the specified period (sampling period) is held as a
judgment value (P.V.).

z

When "Level" is set for "7. Timing input" (page 14)

When the timing input is turned ON, the minimum value of the internal
measurement value (R.V.) during the sampling period is held as the judgment
value (P.V.). When the timing input is turned OFF, the next sampling period begins.

z

When "Edge" is set for "7. Timing input" (page 14)

When the timing input is turned ON, the minimum value of the internal
measurement value (R.V.) during the sampling period is held as the judgment
value (P.V.). Also, when the timing input is turned ON, the next sampling period
begins.

P-H

Display value

Time

Timing Input

Sampling period

Sampling period

Sampling period

ON

OFF

Power ON or
Reset input *

1

After the power is turned on or reset is input, the judgment value (P.V.) is
[- - - - -] until the timing input is first turned ON.

Judgment value (P.V.)
Internal measurement value (R.V.)*

2

Display value

Time

Timing Input

Sampling period

Sampling period

Sampling period

ON

OFF

Power ON or
Reset input *

1

After the power is turned on or reset is input, the judgment value (P.V.) is
[- - - - -] until the timing input is first turned ON.

Judgment value (P.V.)
Internal measurement value (R.V.)*

2

b-H

Display value

Time

Timing Input

Sampling period

Sampling period

Sampling period

ON

OFF

Power ON or
Reset input *

1

After the power is turned on or reset is input, the judgment value (P.V.) is
[- - - - -] until the timing input is first turned ON.

Judgment value (P.V.)
Internal measurement value (R.V.)*

2

Display value

Time

Timing Input

Sampling period

Sampling period

Sampling period

ON

OFF

Power ON or
Reset input *

1

After the power is turned on or reset is input, the judgment value (P.V.) is
[- - - - -] until the timing input is first turned ON.

Judgment value (P.V.)
Internal measurement value (R.V.)*

2

„

Absolute value hold

Within the specified period (sampling period), the absolute value of the maximum
value of the measured value is held as the judgment value (P.V.).

z

When "Level" is set for "7. Timing input" (page 14)

When the timing input is turned ON, the maximum absolute value of the internal
measurement value (R.V.) during the sampling period is held as the judgment
value (P.V.). When the timing input is turned OFF, the next sampling period begins.

z

When "Edge" is set for "7. Timing input" (page 14)

When the timing input is turned ON, the maximum absolute value of the internal
measurement value (R.V.) during the sampling period is held as the judgment
value (P.V.). Also, when the timing input is turned ON, the next sampling period
begins.

„

Auto peak hold

Until the reset input is turned ON, the maximum value is held as the judgment
value (P.V.). External input is not necessary.

„

Auto bottom hold

Until the reset input is turned ON, the minimum value is held as the judgment
value (P.V.). External input is not necessary.

A-H

Absolute value-hold value

Display value

Time

Timing Input ON

0

OFF

Power ON or
Reset input

Sampling period

Sampling period

Sampling period

After the power is turned on or reset is input, the judgment value (P.V.) is
[- - - - -] until the timing input is first turned ON.

Measurement value

Absolute value-hold value

Display value

Time

Timing Input ON

OFF

Power ON or
Reset input

Sampling period

Sampling period

Sampling period

After the power is turned on or reset is input, the judgment value (P.V.) is
[- - - - -] until the timing input is first turned ON.

Measurement value

Auto.P

Auto peak hold value

Display value

Time

Timing Input ON

OFF

Reset input ON

OFF

Measurement value

Auto.b

Auto bottom hold value

Display value

Time

Timing Input ON

OFF

Reset input ON

OFF

Measurement value

Advertising