ManuaL caLiBration

note: Kistler-Morse SVS 2000™ signal processor performs a manual calibration automatically, with Quick Config.

1. Refer to the signal processor manual to determine how to obtain the A/D converter sensitivity, expressed

in Counts/mV. Record this value.

2. Record the Rated Load for one Load Stand.

3. Record the sensitivity (S) for your Load Stand.

Load Stand II

Sensitivity

S2-025K, -075K, -150K, -200K, -300K, -400K,
-500K, -750K, -1M, S2-050K, -100K

26.67 mV/V

4. Record the excitation voltage for the system (typically 12 V).

5. Record the current live load in the vessel.

6. Calculate the Manual calibration values:

Scale Factor Weight= Rated Load (lbs or kg) x Number of supports Scale Factor Counts = S (mV/V) x

excitation Voltage (V) x Counts/mV Zero_Cal = current live load (lbs or kg)

Note: Scale Factor Counts and Scale Factor Weight establish the slope 
of the Manual Calibration Line. Zero_Cal establishes the location of the line.

Indicated

Weight

Weight Signal in Counts

Live Load =

Zero_Cal

Scale Factor Counts

Scale Factor Weight

Scale Factor Counts

Scale Factor Weight

Slope =

Example: You are using four S2-100K (100,000 lb) Load Stand IIs to monitor a vessel.

The vessel currently contains 50,000 lbs of material. The vessel can hold a maximum of 350,000 lbs. Following
the Manual calibration procedure:

1. Counts/mV = 699.05 (from signal processor)

2. Rated load for one Load Stand is 100,000 lbs.

3. S = 26.7 mV/V (from Table 4-1)

4. Excitation voltage = 12 V (from signal processor)

5. Current live load = 50,000 lbs

6. Calculate the calibration values:

Scale Factor Weight

= Rated Load x Number of supports

= 100,000 lbs x 4 legs = 400,000 lbs