9 technical data, Technical data, 9technical data – ALTANA BYK-mac ROBOTIC Manual User Manual

Page 20

Advertising
background image

19

9

Technical Data

Color
Measuring Geometry

Illumination 45°,
-15°,15°, 25°, 45°, 75°, 110° aspecular viewing

Measuring Area

87 x 23 mm

Spectral Range

400 - 700 nm, 10 nm resolution

Measurement Range

0 to 400 % reflectance

Repeatability¹

0.02 dE*
10 consecutive readings on white tile

Reproducibility¹

0.2 dE*
average on 12 BCRA II tiles

Color Scales

dE*; dE CMC; dE 94; dE 2000; dE 99; dE DIN 6175

Illuminants

A; C; D50; D65; F2; F7; F11; F12

Observer

2°; 10°

Effect
Measurement Geometry 15°, 45°, 75° and diffused illumination

perpendicular viewing

Effect Parameters

dS; dS_a; dS_i; dG

Repeatability¹

S_a / S_i: 5% or > 0.50 / G = ± 0.05

Reproducibility¹

S_a / S_i: 10% or > 1.00 / G = ±0.15

Object Curvature

Radius > 400 mm

Measuring Time

< 6 seconds

Memory

1000 Standards/Samples

Language

English, German, French, Italian, Spanish

External Power Supply

Input: 24V ; max. 0.5A

Interface

RS 422

Robotic Requirements

Vibration-free operation

Operating Temperature

10 to 42 °C (50 to 110 °F)

Relative Humidity

up to 85%, 35 °C (95 °F) non-condensing

Dimensions

21 x 12.5 x 17.5 cm (8.3 x 5 x 6.9 in.)

Weight

approx. 3.5 kg (7.7 lbs)

Technical Data

¹ standard deviation

Advertising