Configuring

BYK LC 3

10

Configuring

Note
You can leave the configuration menu at any time with <M>. The parameters
already changed are stored.

Procedure for configuring (factory settings appear in bold typeface):

! Switch off the measuring instrument.
! While pressing the <M> key, press <

>.

! Select the required Baud rate with <

> <>.

Selection: 1200, 2400, 4800, 9600 Baud.

! Then press the <RUN/ENTER> key.
! Set the required calibration interval with <

> <>.

Selection: 1... 180... 999 d.

! Then press the <RUN/ENTER> key.

! Switch the automatic measuring range selection AutoRange on/off with

<

> <>.

Selection: YES (On), no (Off).

! Then press the <RUN/ENTER> key.

! Switchover the reference temperature of the conductivity with <

> <>.

Selection: 25 °C (Tref25) and 20 °C (Tref20).

! Then press the <RUN/ENTER> key.

! Switch the units of the conductivity display on/off with <

> <>.

Selection: S/cm, M

cm.

! Then press the <RUN/ENTER> key.

! Select the date and time step-by-step with <

> <>. In between, press

the <RUN/ENTER> key each time.

! When the last <RUN/ENTER> key has been pressed, the instrument

changes to the last active measuring mode.