A l c, Lc l e – ALTANA LC 3 User Manual
Page 8
Determining/Setting the cell constants
BYK LC 3
8
Procedure for manually setting the cell constant:
Procedure for selecting fixed cell constants 0.010 cm
-1
:
! Press <RUN/ENTER>. The AutoRead measurement begins.
! If the measured value is stable, the instrument displays the determined
cell constants and the calibration evaluation. The measuring instrument
automatically stores the cell constants.
! Switch to the measuring mode with <M>.
3
0 7
4
Cal
cm
1/
Tref25
A L
C
! Press the <CAL> key repeatedly until the CELL display appears.
! Then press the <RUN/ENTER> key.
! Press the <CAL> key until a measured value and a value for the cell con-
stant appears in the required setting range.
! Set the value of the cell constant with <
▲> <▼>.
! Switch to the measuring mode with <M>.
L
C L
E
4
0 1
6
8 0
6
cm
1/
S/
cm
Tref25
nLF
ARng
! Press the <CAL> key repeatedly until the CELL display appears.
! Then press the <RUN/ENTER> key.
! Press the <CAL> key until a measured value and the cell constant
0.010 cm
-1
appear.
! Switch to the measuring mode with <M>.
L
C L
E
0
0 1
0
2 5
9
cm
1/
S/
cm
Tref25
nLF
ARng