Determining/Setting the cell constants

BYK LC 3

8

Procedure for manually setting the cell constant:

Procedure for selecting fixed cell constants 0.010 cm

-1

:

! Press <RUN/ENTER>. The AutoRead measurement begins.

! If the measured value is stable, the instrument displays the determined

cell constants and the calibration evaluation. The measuring instrument
automatically stores the cell constants.

! Switch to the measuring mode with <M>.

3

0 7

4

Cal

cm

1/

Tref25

A L

C

! Press the <CAL> key repeatedly until the CELL display appears.
! Then press the <RUN/ENTER> key.

! Press the <CAL> key until a measured value and a value for the cell con-

stant appears in the required setting range.

! Set the value of the cell constant with <

> <>.

! Switch to the measuring mode with <M>.

L

C L

E

4

0 1

6

8 0

6

cm

1/

S/

cm

Tref25

nLF

ARng

! Press the <CAL> key repeatedly until the CELL display appears.
! Then press the <RUN/ENTER> key.

! Press the <CAL> key until a measured value and the cell constant

0.010 cm

-1

appear.

! Switch to the measuring mode with <M>.