Determining/setting the cell constants, A l c, Lc l e 2 0 7 4 – ALTANA LC 3 User Manual

Page 7

Advertising
background image

BYK LC 3

Determining/Setting the cell constants

7

Determining/Setting the cell constants

Procedure

The cell constants are determined in the control standard, 0.01 mol/l KCl.

You can determine the cell constants of the conductivity measuring cells in
the range of 0.450 ... 0.500 cm

-1

or 0.800 ... 1.200 cm

-1

.

You can also set the cell constants manually in the ranges

! 0.090 ... 0.110 cm

-1

! 0.250 ... 2.500 cm

-1

or select the fixed cell constants 0.010 cm

-1

.

Calibration

evaluation

After the calibration, the measuring instrument automatically evaluates the
current status of the calibration. The evaluation appears on the display.

Procedure for determining the cell constants:

Display

Cell constant [cm

-1

]

0.450 ... 0.500 cm

-1

0.800 ... 1.200 cm

-1

E3

Invalid calibration

Outside the range
0.450 ... 0.500 cm

-1

or
0.800 ... 1.200 cm

-1

! Connect a conductivity measuring cell to the measuring instrument.
! Press the <CAL> key repeatedly until the CELL display appears.
! Then press the <RUN/ENTER> key.

! Press the <CAL> key repeatedly until the CAL appears. Depending

on the calibration status, the instrument displays:

– the current, calibrated cell constant (with sensor symbol on the

display) or

– the fixed cell constant 0.475 1/cm (without sensor symbol on the

display). In this case, the measurement parameters are initialized.

Note
At this point, the procedure can be terminated with <M>.

! Immerse the conductivity measuring cell in the control standard solution,

0.01 mol/l KCI.

L

C L

E

2

0 7

4

Cal

cm

1/

Tref25

A L

C