12. TECHNICAL SPECIFICATIONS

Weight: 1.07 lbs. (487 g) (6021N, 6022N with batteries,

850-6022N soft cuff)
1.14 lbs. (515 g) (6023N with batteries,
850-6023N rigid cuff)

Size: 5.35” x 3.78” x 2.28”

136mm x 96mm x 58mm (including cuff)

Storage temperature: 23°F−122°F (–5°C− +50°C)

Humidity: 15 to 85% relative humidity maximum

Operation temperature: 50°F−104°F (10°C−40°C)

Display: LCD (Liquid Crystal Display)

Measuring method: Oscillometric

Pressure sensor: Capacitive

Measuring ranges

SYS/DIA Pressure: SYS/DIA: 30 to 280 mmHg

Pulse: 40 to 200 beats per minute

Cuff pressure display range: 0-299 mmHg

Memory: Automatically stores the last 99 measurements

for 2 users (198 total).

Measuring resolution: 1 mmHg

Accuracy: Pressure within ± 3 mmHg or 2% of the reading

≥200mmHg

Pulse ± 5 % of the reading

Power Source: 4 “AA” batteriesor 1.5V AC adapter

Includes: Automatic Blood Pressure Monitor and Cuff, 4

AA batteries, Instruction Book, and Record Book.
Other cuffs are also available as a special accessory:

Item No

Size

Range

850-6021NSA Sm. Adult 6.7−8.7” (17−22 cm)
850-6022N

Adult

8.7−16.5” (22−42 cm)

850-6022NX

Lg. Adult

12.5−20.4” (32−52 cm)

850-6023N

Adult

8.7−18.1” (22−46 cm)

13. WARRANTY

This blood pressure monitor is warranted for 5 years from date of purchase. This
warranty includes the instrument and the cuff. The warranty does not apply to damage
caused by improper handling, accidents, improper use, or alterations made to the
instrument by third parties. The warranty is only valid after the product is registered online
at www.adctoday.com.

14. QUALITY STANDARDS

Device standard:

This device is manufactured to meet the European and United States standards for non-
invasive blood pressure monitors:

EN1060-1 / 1995 • EN1060-3 / 1997 • EN1060-4 / 2004

Advertising