9 trouble shooting – Envirco MAC 10 Original User Manual

Page 18

Advertising
background image

MAC 10 ORIGINAL VE5 SERVICE MANUAL

PAGE 17

9 Trouble Shooting:

Low Air Velocity


Step 1. Check prefilter media; replace or clean as necessary.

Step 2. Flip switch from “Low” to “Medium” or “Medium” to “High” on units with 3-speed

switch.


Step 3. Adjust variable speed control (only units with optional Speed Control) for higher

blower output.


Step 4. Check power supply for proper voltage, amperage and distribution frequency.

Step 5. Replace HEPA/ULPA filter if the air velocity remains low.


High Air Velocity


Step 1.

Flip switch from “HIGH” to “MEDUIM” to “LOW” on units with 3-speed switch.

Step 2. Adjust variable speed control (only units with optional Speed Control) for lower

blower output.




Non-Laminar Flow and/or Excessive Contamination

Step 1. Insure that no large obstructions are upstream of airflow pattern.

Step 2. Determine that no other air-moving devices are operating in or around clean

room which disrupt room’s airflow pattern.


Step 3. Check air velocity and if low, conduct the “Low Air Velocity” procedure outlined

above.

Step 4.

Conduct smoke and photometer test on HEPA filter. Seal or replace HEPA filter
as necessary.


Advertising