KERN & SOHN OEX Safety User Manual

Page 10

Advertising
background image

10

OEX-BA-e-1410

CAUTION!

Installing of the indicator, placing the weighing platform and

connecting grounding must be carried out at a time when there
is no risk of the emergence of an explosive atmosphere.

Place the weighing platform on an even and stable foundation distant from

source of heat, drafts and leveled.

Level the scale using adjustable feet and a level.

Check whether the adjustable feet rest on the scale’s foundation.

Each adjustable feet should take an equal amount of load.

Double check the level of scale location. In correct leveling, the air bubble should

be located within the centre of level’s ring.

If the weighing platform requires leveling, then screw scale’s adjustable feet to
adjust level.

Check the grounding-mass springs between the weighing platform and the

weighing pan ensuring the discharge of static electricity from the weighing pan.

Put the weighing pan onto the weighing platform,

Place the power adapter outside the explosive zone (i.e. in a safe area),

Plug the cable (~20 m) to the power outlet DC IN

on scale’s indicator

KEX-TM.


Advertising