12

OEX-BA-e-1410

6 Changing the scale place of operation

While moving a scale in another place of operation, all safety

measures listed in point 5 of this instruction should be taken.

7 Cleaning and maintenance

Unplug the power adapter IEX-A01 from mains.

Check wether the indicator´s electrostatic grounding is correctly plugged and

is not defected.

Clean the scale and the housing with a slightly moistened cloth. You

can use typical domestic cleaning agents.

For clearing an overlay of the indicator do not use agents that can

mechanically scratch the surface. It includes an internal conductive
layer that is important for save operation.

Platform levelling feet and shock absorbers are made of rubber.

Within the hazardous area they need to be cleaned with a slightly
moistened cloth. Moisture protects against gathering electrostatic
charges.

While cleaning the platform do not damage the masking springs.

They are for gathering electrostatic charges from the weight pan.
They should be ~5mm above the shock absorbers.

Cleaning may not be performed when there are explosive gases.

There is forbidden to us concentrated acids, bases, solvents or

alcohol for cleaning the device.

There is also forbidden to use compressed air and high pressure

washing equipment.

8 Technical surveys

Technical inspection of a scale operating in explosive area has to be

carried out by a trained personnel at least once every three months.


Each technical inspection has to cover:

Condition of scale’s indicator electrostatic grounding, including:

o

Its mounting to the indicator’s housing and connection resistance - max.
100Ω,

o Mounting of a steel flexible tube protecting a conductor to the weighing

platform and the indicator (reg. scal

es with indicator “on a cable”).

Quality of the connection between the platform and the indicator (pliant steel pipe

connecting the two subassemblies).

The damage and loosing one of connection is inadmissible.

Electrical connection between the housing and the platform - max. resistance

between the weight pan (put on) and the housing should be lower than 100Ω.

Stability and levelling of the scale.