CAUTION!

In order to dissipate static electricity, it is necessary to ground the

scale’s indicator.

Places for plugging the wires are marked with “ ” symbol.


Before moving the scale to its intended place of use, prepare a conductor for

electrostatic grounding of the indicator.

The grounding conductor should be ended with a loop that is screwed to the

indicator using M4 nut.

In order to ensure constant holding down the loop to the indicator´s housing, use

tooth lock washer.

The grounding conductor should have yellow and green colour with minimal

intersection 4mm².

Below drawing presents the means of scale grounding:


Place the weighing platform and the indicator in its intended place of use in the

EX zone and ground them.


Transport protections

Weighing pan